Sec. 30-35. Sterilization.
About this text
- As published by Municode, current through 2023-04-18
- The enacted text only: the publisher's notes (editor's notes, state law references, cross references) are not included.
(a) Instrument cleaning.
(1) All non-single-use, non-disposable instruments shall be sterilized after each use according to the manufacturer's recommendations.
(2) Body craft studio employees shall wear heavy-duty, multi-use, waterproof gloves while cleaning instruments.
(3) Used instruments shall be soaked in an enzymatic solution or cool water until cleaning can be performed.
(4) Instruments shall be disassembled for cleaning.
(5) All instrument components shall be cleaned, either manually or in an ultrasonic cleaner, using the appropriate cleaning agent specific to the type of cleaning performed.
(b) Instrument packaging.
(1) Body craft studio employees shall wear clean gloves while packaging or wrapping instruments.
(2) Instruments shall be wrapped or packaged with a sterilizer indicator on each package.
(3) All packages shall be labeled with the time and date of sterilization. Packages shall no longer be considered sterile if the integrity of the package is damaged.
(c) Instrument sterilization.
(1) Except as provided in this subsection, all body craft studios shall utilize a sterilizer that shall be designed and labeled as a medical instrument sterilizer to achieve the sterilization definition.
(2) The operator's manual for the sterilizer shall be available on the premises at all times and the sterilizer shall be operated according to the manufacturer's recommendations.
(3) The sterilizer shall be cleaned and maintained according to the manufacturer's recommendations.
(4) A sterilizer load log shall be maintained according to the manufacturer's recommendations for the life of the facility and made available for inspection by the director and the customer or client. The log shall contain the following documentation for each load:
a. A description of instruments contained in the load;
b. The date of sterilization, and time if more than one load is processed in a single day;
c. Manufacturer's recommended cycle data;
d. An indication of the proper sterilization of instruments, as indicated by the appropriate color change of the sterilizer indicator on each package. The indicator used shall be compatible with the sterilization process being used;
e. Any action taken when appropriate color indicator change did not occur;
f. Maintenance log, including:
1. Date of service;
2. Sterilizer model and serial number;
3. Sterilizer location;
4. Name of person requesting service;
5. Description of malfunctions;
6. Name of person performing maintenance;
7. Description of services and parts replaced;
8. Results of biologic validation and testing;
9. Signature and title of person acknowledging completed work.
(d) Sterilization monitoring.
(1) Sterilizer monitoring shall be performed at least monthly, unless the manufacturer specifies more frequent monitoring, or after 40 hours of usage, whichever comes first, by using a commercial biological monitoring (spore) system.
(2) All biological indicators shall be analyzed by a laboratory that operates under FDA guidelines and is independent of the body craft studio, with reports sent to the studio.
(3) Biological indicator test results shall be maintained on the premises for the life of the studio and must be available for inspection by the director at all times.
(4) Upon notification of a positive microbiological monitoring report, the sterilizer shall be immediately checked for proper use and function and the owner shall cease use of the sterilizer immediately upon receipt of the positive report. Three consecutive passed biological tests must be achieved before the studio is reopened. The studio shall have the option to obtain a properly functioning sterilizer in order to remain open.
(e) Instrument storage.
(1) Hands shall be washed and gloved prior to handling sterilized instruments.
(2) After sterilization, the instruments shall be stored in a dry, clean area reserved for storage of sterilized instruments.