Ir al contenido
Georgia Commons
Menú

Columbia County · Código de ordenanzas

Part II - CODE OF ORDINANCES / Chapter 391-3-4.15

Sec. 46-126. Sterilization.

Sobre este texto

  • As published by Municode, current through 2026-02-17
  • The enacted text only: the publisher's notes (editor's notes, state law references, cross references) are not included.

Las notas de arriba, el texto de la sección y su línea de historia se muestran en inglés, tal como los publica la fuente y como los da el corpus.

(a) An ultrasonic cleaning unit and operational medical grade autoclave shall be provided in

each body art studio.

(b) Ultrasonic cleaning units shall be clearly labeled "biohazardous" and placed away from

the autoclave and workstations.

(c) The ultrasonic cleaning unit and medical grade autoclave shall be used and maintained according to manufacturer's specifications. Each ultrasonic cleaning unit and medical grade

autoclave shall be emptied and thoroughly cleaned and disinfected at least once each day that the unit is used. The ultrasonic cleaning unit and autoclave shall be made available for inspection.

(d) Used non-disposable instruments shall be kept in a separate puncture resistant container until cleaned. The cleaning method shall include the following:

(1) Instruments shall be placed into an ultrasonic cleaning unit

(2) After removal from the ultrasonic cleaning unit, instruments shall then be brush scrubbed in hot water and soap, then

rinsed in clean water

(3) Instruments shall then be packed individually in sterilized packs and sterilized in a medical grade autoclave. All sterilized packs shall contain either a

sterilized indicator or internal temperature indicator.

(4) Prior to being placed in the autoclave all equipment shall be bagged, labeled,

initialed, dated and sealed.

(5) Each autoclave bag may hold no more than one individual item (ex. one tube or needle bar).

(e) After sterilization, the packaged instruments shall be stored in a clean dry cabinet or other tightly covered container reserved and

labeled for storage of sterile instruments.

(f) If a sterilized package has been breached or allowed to get wet, the instrument(s) must be re-packaged and sterilized again before use.

(g) A cleaning schedule log of the sterilization equipment must be maintained at the body art

establishment.

(h) Spore indicators shall be used a minimum of at least once a month and the results must be kept at the body art establishment for a minimum of two years. An independent commercial testing laboratory contracted by the operator/owner shall perform monthly biological spore testing of the autoclave. A provision shall be included in the contract between the operator/owner with the

commercial testing laboratory requiring the commercial testing facility to notify the board of health of any failure of the autoclave to eradicate all living organisms, including spores.

(i) Upon notification of a positive microbiological monitoring report, the sterilizer shall be immediately checked for proper use and function and the operator/owner shall cease use of the

sterilizer immediately upon receipt of the positive report. Three consecutive negative biological tests must be achieved before the body art establishment is re-opened. The body art

establishment shall have the option to obtain a properly functioning sterilizer with a negative biological report and to remove from the premises the sterilizer with the positive report in order to remain open.

Historia

La línea de historia que la fuente imprime al final de la sección.

(Ord. No. 11-12, 3-15-2011)

Leer este código donde se publica (Municode) La fuente es el código del condado tal como lo publica su editorial o su propio sitio web. Consúltelo allí para ver cambios posteriores.

Versión local-code:columbia-county/code/46-126@2026-02-17-af31eba9a3

El asistente puede decir lo que dice esta sección en esta edición, citándola y enlazándola. No aplicará la ley a su situación, y nada de esto es asesoría legal.